Filter Integrity Tester, Bubble Point Tester, Audit Optional

Filter Integrity Tester, Bubble Point Tester, Audit Optional

FeaturesPowerful functions, covering all the existing test methods for the filter integrity testing; The first domestic
Basic Info.
Model NO. FIT-6.5
Operation Pressure 100-10000 Mbar (1.5-145psi) ;
Voltage 100 – 260V/ 50Hz
Transport Package Export Packing
Specification 405*240*270 (with handle) 405*240*220 (without han
Trademark Xiangyi
Origin China
Production Capacity 100 PCS/Month
Product Description
FeaturesPowerful functions, covering all the existing test methods for the filter integrity testing;
The first domestic implementation of integrity testing for the ultrafiltration system;
Scientific authority management design, password, permissions, authority grading, electronic signature, etc., in full compliance with FDA 21 CFR PART 11 requirements;
It can create120 groups of reservation solution, so the instrument is more simplistic and intelligent;
With rich data interface, the instrument not only includes the standard configuration of digital and analog interface (RS232/USB), but also can customize a variety of industrial bus and simulation control port according to customer's need, in preparation for the intelligent system upgrade, such as the complete set of online automatic sterilization control system;
Independent research and development team can carry out the design of specific solutions according to the needs of customers; Many years of field experience and strong professional services can not only ensure that the customers use the instrument at ease, but also can provide the technical support for the design and configuration of the filter system of the customers;
Specifications
ModelFIT-6.5
Voltage100 - 260V/ 50HZ
Power120W
Operation Pressure100-10000 mbar (1.5-145psi) ;
Pressure Unitmbar
Ambient Temperature+5ºC ~ +40ºC
Relative Humidity10-80%
Dimension (mm) (D*W*H)405*240*270 (with handle)
405*240*220 (without handle)
FunctionManual Bubble point Test;
Basic Bubble Point Test;
Extensive Bubble Point Test;
Pressure Holding Test;
Diffusion FLow Test;
Water Intrusion Test;
Ultrafitration Membrane Test
Test AccuracyUpstream Volume Test: ≤± 4%
Bubble Point Test: ≤± 50mbar
Diffusion Flow Test: ≤ ± 4%
Water Immersion: ≤± 0.01ml
Test ScopeBP: 100-8000mbar
DF: 1-600ml/min
WI: 0.01-100ml/min
Applied RangeSymmetric and asymmetric membrane test
needle filter
filter bag
flat filter
filter cartridge (big filter volume )
ultrafiltration membrane
ultrafiltration column
a variety of irregular filters
Authority ManagementUser Name and Password to Login
Meeting FDA 21CFR PART 11
Reservation Solution120 sets
Audit trail (customization)More subject logs
Print FunctionBuilt-in printer
Historical RecordsMore than1 5000 sets unlimited sets records
Records CopyUSB(Include test curve)
Display Screen7'' color touch screen
Serial ConnectionRS232/USB
LanguageEnglish
Work ModeOnline/Offline
Other functionsOne/two-dimension cods , Scanning gun; Lithium battery (customization)
Applicable EnvironmentAbove D level
Weight8KG
Anti-liquid levelFront IP65
Signal output4-20mA
RS485
12V Alarm output