Spectrometer Metal Analysis Laboratory Instrument

Spectrometer Metal Analysis Laboratory Instrument

Specification:Metal analyzerCCD-based, spark discharge, optical emission spectrometerSoftware-extendable configuration f
Basic Info.
Model NO. W6
Analytical Method Full Digital and Spectrum Direct Reading
CCD Pixel Resolution 6pm
Transport Package Wooden Package
Trademark wuxi jinyibo
Origin Jiangsu Wuxi
HS Code 9027300000
Production Capacity 30 Sets/Month
Product Description
Specification:Metal analyzerCCD-based, spark discharge, optical emission spectrometerSoftware-extendable configuration fro utmost elemental flexibilityUltra-low limits of detectionUltra-high speed of measurementLong-term stability and repeatabilityExcellent vertical optical emission spectroscopySummary:W6-The most advanced Arc/Spark-OES with Ultimate Performance for Metal Analysis. It use full-digital technology to replace bulky photomultiplier tube (PMT) simulation technology and in the lead of international spectrometer technology. The adoption of vacuum optical chamber design, full-digital excitation light source, advanced CCD detectors, and high-speed data readout system equips the device with high properties, ultra-low limit of detection (LOD), long-term stability and repeatability. Its analytical precision reaches the world leading level. It is one of the major devices that applies to quality control, material brand recognition, material research and development in the metallic manufacturing, processing and smelting industry. This analyzer sets a new standard for R&D and process/quality control.Applications:W6 optical emission spectrometers(spark OES) as one of the major devices that applies to quality control, material brand recognition, material research and development in the metallic manufacturing, processing and smelting industry.W6 are used for applications in metal elemental analysis, trace element analysis for science and industry like Metallurgy, Foundry, Mechanical Engineering, Scientific Research, Product Inspection, Automobile, Petrochemical Engineering, Shipbuilding, Electrics, Aerospace, Nuclear Power, Metallic and Nonferrous Metallic Smelting, Processing and Recycling.High-end Laboratories Defence, Railways, pure research, etc.Pure metal applications 99.9%+ purity Al, Pb, Zn, Mg etc.Regulatory compliance Very low LODs to control Pb, Cd, As etc. (e.g. ship-making)Integrated Steel Plants C, S, P, B, N, Ti;Manufacturing FacilitiesWarehouse material identificationBase: Fe, Cu, Al, Ni, Co, Mg, Ti, Zn, Pb, Sn, Ag etcIron(Fe) and its alloys (Steel alloy, Cast iron, Fe-Low alloy, Fe-Cr steel, Fe-Cast iron, Fe-Cr-cast, Fe-Mn steel, Fe-Tool steel etc.)Aluminium(Al) and its alloys (Al-Si alloy, Al-Zn alloy, Al-Cu alloy, Al-Mg alloy, Pure-Al alloy etc)Copper(Cu) and its alloys (Brass, Copper-Nickel-Zn, Aluminum Bronze, Tin-lead Bronze, Red Copper, Be-Bronze, Si-Bronze etc)Nickel(Ni) and its alloys (Pure Ni, Monel metal, Hadtelloy Alloy, Incoloy, Inconel, Nimonic etc)Cobalt(Co) and its alloys (Co-orientation, Low Co alloy, Stellite 6,25,31, Stellite 8,WI 52, Stellite 188, F)Magnesium(Mg) and its alloys (Pure Mg, Mg/Al/Mn/Zn-alloys)Titanium(Ti) and its alloysZinc(Zn) and its alloysLead(Pb) and its alloysTin(Sn) and its alloysArgentum(Ag) and its alloysSmall sample, special size sample and wire detectionParameter:
Item Index
Optical System Focal Length400nm
Wavelength range130nm-800nm
Detector High resolution CCD Multi detectors
Degree of vacuumAuto control within 6-20 pa
Pixel resolution30pm
Grating line2400m1/mm
First order spectral line dispersion rare1.2nm/mm
Average resolution ratio10pm/pixel
Full spectrum
Light room temperature is controlled automatically
Spark SourceType Digital arc and spark source
Spark frequency100-1000HZ
Plasma current 1-80A
Ignition voltage>15000V

Excitation light
Optimization of discharge parameters design
High energy precombustion technology HEPS

Processor
High-speed data synchronization acquisition and processing
Spark Stand Electrode Tungsten electrode technology
Make up Thermal deformation self-compensation design
Argon flushed with minimal consumption of Argon
Spray discharge electrode technology
Adjustable electrode technology
OthersMeasurable elementsFe,Al,Cu,Ni,Ti,Co,Zn,Sn,Mg,Pb etc
Dimension 650mm(L)*860mm*1200mm(H)
weightAbout 235kg
Storage temperature0ºC-45ºC
Operating temperature10ºC-35ºC, 23±2ºC is recommended
Power AC220V/50Hz(Customized)
Power consumptionExcitation:700W/Stand by:100W
Argon quality99.999%, Argon pressure>4Mpa
Argon consumption 5L/min during spark mode
Interface Ethernet data transmission based on DM9000A