Syv4.0 Bubble Point Instrument Filter Integrity Tester

Syv4.0 Bubble Point Instrument Filter Integrity Tester

Filter Integrity Tester IntroductionIntegrity Tester are designed for testing integrity of filters and filter systems .
Basic Info.
Model NO. SY-V4.0
Origin China
HS Code 9031809090
Production Capacity 50000ses/Year
Product Description

Filter Integrity Tester

IntroductionIntegrity Tester are designed for testing integrity of filters and filter systems . The State Pharmacopoeia and GMP specification requirements. The V4.0 Integrity Tester is compact, handy to use, and fully automated integrity test instrument, which performs bubble point, diffusion flow, enhanced bubble point, and water-based test for hydrophobic filters. The first domestic launch of integrity test for water-based test for hydrophobic filters to meet different customer needs.Specifications
Power Supply100 - 260 VAC, 50HZ, 120W ;
Operation pressure100-8000mbar (115psi) ;
Unitmbar
Operation ConditionAmbient Temperature:+5ºC ~ +40ºC;Relative Humidity:10-80%
Dimension(mm)400(Depth)*240(Width)*270(Height1)*220(Height2)
Test FunctionManual Bubble point Test;Basic Bubble Point Test;Extensive Bubble Point Test;Diffusion Flow Test;Water Immersion Test;
Test AccuracyUpstream Volume:± 4%;Bubble Point:≤± 50mbar ;Diffusion Flow Test:≤± 4% ;Water Immersion Test:≤± 0.05ml
Test ScopeBubble Point :590-6900mbar Diffusion Flow Speed:1-180ml/min WI:0.01-30ml/min
Applied RangeSymmetric and asymmetric membrane test, needle filter, filter bag, flat filter, filter cartridge (3 core within 20 inches)
Print FunctionBuilt-in printer;
Historical Records500 sets;
Display Screen5.7″TFT,Single color;
Serial ConnectionRS232;
LanguageChinese/English;
Work ModeOnline/Offline;
Applicable EnvironmentAbove B level;
Weight7.5KG;

Main features
1.The scientific electronic signature and classification to distinguish the responsibility easily, and to prevent false operation.2.Display of test data and curves in real time, monitoring the whole process of testing.3.Provides automatic printing set function, the user can operate more simple and convenient.4.Water Based test (WH) for Hydrophobic Filter: using purified water as test liquid instead of IPA and Ethanol, thus it can avoid trace quantity of ethanol or IPA contamination when testing filters.5.The instrument can store 500 history record and curve.
Applied range
Disc Membrane:Φ25mm-Φ300mm;Standard Cartridge:2.5″- 40″Capsule & Mini CatridgeAir filter test :2.5″- 40